Voltage Range: Two ranges: 0V to 50V or 0V to 80V
Current Range: 0.01A to 50A
Power Range: Up to 750W: within envelope shown (12V/50A, 36V/20A, 60V/12A & 80V/9.4A guaranteed). Or programmable power limit 100W to 600W in 10W increments.
Voltage Setting: Resolution: 1mV (50V Range), 2mV (80V Range)
Accuracy: 0.1% of setting ± 2mV, (± 4mV on 80V range) Current Setting: Resolution 10mA Accuracy: 0.3% of setting ± 20mA
Operating Mode: Constant voltage, constant current or constant power with automatic cross-over dependant on output setting parameters. CV, CI, or CP mode indication in display.
Output Switch: Independent electronic switching, with ON indication. Pre-set voltage and current displayed when off.
Output Terminals: Front & rear panel safety terminals accepting 6mm wire diameter, 6mm plugs or 8mm spades at 50 Amps max, or 4mm plugs at 30 Amps max.
Sensing: Switchable between local and remote. Screwless remote sense terminals on both front & rear panels. Sense miswiring trip and indication.
Ripple & Noise (20MHz bandwidth): Typically, <3mVrms, <20mV pk-pk at maximum load, CV mode
Load Regulation: Change in output for any load change within PowerFlex envelope, using remote sense: Constant voltage: <0.01% ± 5mV
Line Regulation: Change in output for a 10% line change: Constant voltage: <0.01% ± 5mV
Transient Response: To within 100mV of set level for a 5% to 95% load change. Rear Terminals: <2ms Front Terminals: <2.5ms
Output Protection: Output will withstand an applied forward voltage of up to 90V. Reverse protection by diode clamp for reverse currents up to 3A.
Over-voltage Protection (OVP): Range 2V to 90V. Resolution 0.1V; accuracy: 0.2% ± 0.2V.Response time typically 100µs.
Over-current Protection (OCP): Range 2A to 55A. Resolution 0.1A; accuracy: 0.5% ± 0.2A. Response time typically 100ms.
Over-temperature Protection: The output will be tripped off if a fault causes the internal temperature to rise excessively.
Temperature Coefficient: Typically, <100ppm/°C